Arid
DOI10.1002/bmc.5128
Sensitive, wide-range and high-throughput quantification of cyclosporine in whole blood using ultra-performance liquid chromatography coupled to tandem mass spectrometry and comparison with an antibody-conjugated magnetic immunoassay
Watanabe, Takuma; Tanaka, Ryota; Ono, Hiroyuki; Suzuki, Yosuke; Tatsuta, Ryosuke; Itoh, Hiroki
通讯作者Tanaka, R (corresponding author), Oita Univ Hosp, Dept Clin Pharm, Hasama Machi, Yufu, Oita 8795593, Japan.
来源期刊BIOMEDICAL CHROMATOGRAPHY
ISSN0269-3879
EISSN1099-0801
出版年2021
卷号35期号:8
英文摘要Because either trough or peak concentration at 2 h after administration is measured in routine therapeutic drug monitoring for cyclosporine A (CyA), a quantification method with a wide-range calibration curve capable of simultaneously measuring both concentrations is required. We developed a sensitive, wide-range and high-throughput quantification method for CyA in whole blood using ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS), and compared patients' blood CyA levels measured by UPLC-MS/MS and antibody-conjugated magnetic immunoassay (ACMIA). Whole blood samples were prepared by solid-phase extraction using Oasis HLB mu Elution plate. The UPLC-MS/MS assay showed excellent linearity over a wide calibration range of 5-2500 ng/mL. Within-batch accuracy and precision as well as batch-to-batch accuracy and precision fulfilled the criteria of US Food and Drug Administration guidelines. The blood CyA concentrations measured by the UPLC-MS/MS assay correlated strongly with those measured by ACMIA. A Bland-Altman plot showed a fixed error between CyA concentrations measured by the two methods, and the concentrations measured by the UPLC-MS/MS method were consistently lower than those measured by ACMIA. We have succeeded to develop a sensitive, wide-range and high-throughput quantification method for CyA in whole blood using UPLC-MS/MS.
英文关键词antibody‐ conjugated magnetic immunoassay cyclosporine therapeutic drug monitoring ultra‐ performance liquid chromatography– tandem mass spectrometry
类型Article
语种英语
收录类别SCI-E
WOS记录号WOS:000641125700001
WOS类目Biochemical Research Methods ; Biochemistry & Molecular Biology ; Chemistry, Analytical ; Pharmacology & Pharmacy
WOS研究方向Biochemistry & Molecular Biology ; Chemistry ; Pharmacology & Pharmacy
资源类型期刊论文
条目标识符http://119.78.100.177/qdio/handle/2XILL650/349716
作者单位[Watanabe, Takuma; Tanaka, Ryota; Ono, Hiroyuki; Tatsuta, Ryosuke; Itoh, Hiroki] Oita Univ Hosp, Dept Clin Pharm, Hasama Machi, Yufu, Oita 8795593, Japan; [Suzuki, Yosuke] Meiji Pharmaceut Univ, Dept Medicat Use Anal & Clin Res, Tokyo, Japan
推荐引用方式
GB/T 7714
Watanabe, Takuma,Tanaka, Ryota,Ono, Hiroyuki,et al. Sensitive, wide-range and high-throughput quantification of cyclosporine in whole blood using ultra-performance liquid chromatography coupled to tandem mass spectrometry and comparison with an antibody-conjugated magnetic immunoassay[J],2021,35(8).
APA Watanabe, Takuma,Tanaka, Ryota,Ono, Hiroyuki,Suzuki, Yosuke,Tatsuta, Ryosuke,&Itoh, Hiroki.(2021).Sensitive, wide-range and high-throughput quantification of cyclosporine in whole blood using ultra-performance liquid chromatography coupled to tandem mass spectrometry and comparison with an antibody-conjugated magnetic immunoassay.BIOMEDICAL CHROMATOGRAPHY,35(8).
MLA Watanabe, Takuma,et al."Sensitive, wide-range and high-throughput quantification of cyclosporine in whole blood using ultra-performance liquid chromatography coupled to tandem mass spectrometry and comparison with an antibody-conjugated magnetic immunoassay".BIOMEDICAL CHROMATOGRAPHY 35.8(2021).
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Watanabe, Takuma]的文章
[Tanaka, Ryota]的文章
[Ono, Hiroyuki]的文章
百度学术
百度学术中相似的文章
[Watanabe, Takuma]的文章
[Tanaka, Ryota]的文章
[Ono, Hiroyuki]的文章
必应学术
必应学术中相似的文章
[Watanabe, Takuma]的文章
[Tanaka, Ryota]的文章
[Ono, Hiroyuki]的文章
相关权益政策
暂无数据
收藏/分享

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。