Knowledge Resource Center for Ecological Environment in Arid Area
DOI | 10.1016/j.talanta.2016.06.032 |
Pre-analytical and analytical validations and clinical applications of a miniaturized, simple and cost-effective solid phase extraction combined with LC-MS/MS for the simultaneous determination of catecholamines and metanephrines in spot urine samples | |
Li, Xiaoguang (Sunny)1; Li, Shu1; Kellermann, Gottfried1,2 | |
通讯作者 | Li, Xiaoguang (Sunny) |
来源期刊 | TALANTA
![]() |
ISSN | 0039-9140 |
EISSN | 1873-3573 |
出版年 | 2016 |
卷号 | 159页码:238-247 |
英文摘要 | It remains a challenge to simultaneously quantify catecholamines and metanephrines in a simple, sensitive and cost-effective manner due to pre-analytical and analytical constraints. Herein, we describe such a method consisting of a miniaturized sample preparation and selective LC-MS/MS detection by the use of second morning spot urine samples. Ten microliters of second morning urine sample were subjected to solid phase extraction on an Oasis HLB microplate upon complexation with phenylboronic acid. The analytes were well-resolved on a Luna PFP column followed by tandem mass spectrometric detection. Full validation and suitability of spot urine sampling and biological variation were investigated. The extraction recovery and matrix effect are 74.1-97.3% and 84.1-119.0%, respectively. The linearity range is 2.5-500, 0.5-500, 2.5-1250, 2.5-1250 and 0.5-1250 ng/mL for norepinephrine, epinephrine, dopamine, normetanephrine and metanephrine, respectively. The intra- and inter-assay imprecisions are <= 9.4% for spiked quality control samples, and the respective recoveries are 97.2-112.5% and 95.9-104.0%. The Deming regression slope is 0.90-1.08, and the mean Bland-Altman percentage difference is from -3.29 to 11.85 between a published and proposed method (n=50). A correlation observed for the spot and 24 h urine collections is significant (n=20, p < 0.0001, r: 0.84-0.95, slope: 0.61-0.98). No statistical differences are found in day-to-day biological variability (n=20). Reference intervals are established for an apparently healthy population (n=88). The developed method, being practical, sensitive, reliable and cost-effective, is expected to set a new stage for routine testing, basic research and clinical applications. (C) 2016 Elsevier B.V. All rights reserved. |
英文关键词 | Catecholamine Metanephrine LC-MS/MS Miniaturized solid phase extraction (SPE) Phenylboronic acid (PBA) Spot urine |
类型 | Article |
语种 | 英语 |
国家 | USA |
收录类别 | SCI-E |
WOS记录号 | WOS:000381322000034 |
WOS关键词 | TANDEM MASS-SPECTROMETRY ; PERFORMANCE LIQUID-CHROMATOGRAPHY ; PLASMA-CATECHOLAMINES ; BIOLOGICAL VARIATION ; ALZHEIMERS-DISEASE ; PHEOCHROMOCYTOMA ; METABOLITES ; NEUROTRANSMITTERS ; NORMETANEPHRINE ; QUANTIFICATION |
WOS类目 | Chemistry, Analytical |
WOS研究方向 | Chemistry |
资源类型 | 期刊论文 |
条目标识符 | http://119.78.100.177/qdio/handle/2XILL650/196624 |
作者单位 | 1.Pharmasan Labs Inc, 373 280th St, Osceola, WI 54020 USA; 2.NeuroScience Inc, 373 280th St, Osceola, WI 54020 USA |
推荐引用方式 GB/T 7714 | Li, Xiaoguang ,Li, Shu,Kellermann, Gottfried. Pre-analytical and analytical validations and clinical applications of a miniaturized, simple and cost-effective solid phase extraction combined with LC-MS/MS for the simultaneous determination of catecholamines and metanephrines in spot urine samples[J],2016,159:238-247. |
APA | Li, Xiaoguang ,Li, Shu,&Kellermann, Gottfried.(2016).Pre-analytical and analytical validations and clinical applications of a miniaturized, simple and cost-effective solid phase extraction combined with LC-MS/MS for the simultaneous determination of catecholamines and metanephrines in spot urine samples.TALANTA,159,238-247. |
MLA | Li, Xiaoguang ,et al."Pre-analytical and analytical validations and clinical applications of a miniaturized, simple and cost-effective solid phase extraction combined with LC-MS/MS for the simultaneous determination of catecholamines and metanephrines in spot urine samples".TALANTA 159(2016):238-247. |