Arid
DOI10.1177/0310057X0603400112
The performance and safety of a pleural drainage unit under hyperbaric conditions
Walker, KJ; Millar, IL; Fock, A
通讯作者Walker, KJ
来源期刊ANAESTHESIA AND INTENSIVE CARE
ISSN0310-057X
EISSN1448-0271
出版年2006
卷号34期号:1页码:61-67
英文摘要

The performance of a proprietary dry suction pleural drainage unit was measured under hyperbaric oxygenation conditions. The test pleural drainage unit was connected to pressure gauges that allowed the pressures created in the suction and collection chambers to be measured as well as the pleural drainage catheter pressures under varied suction regulator settings during compression, hyperbaric steady states and decompression. The maximum flow capacity of the unit was also measured under varying hyperbaric conditions.


The Atrium Oasis Dry Suction 3600 Chest Drain brand was dramatically affected by pressure change. Nevertheless, based upon our testing, we believe it call be used safely in a hyperbaric environment provided that the following precautions are taken.


Suction should not be applied during pressurization. Pressurization needs to be slow, 10 kpa/min or less. Suction is needed for air leaks of 4/min or more at pressure. At stable hyperbaric pressure, the level of suction delivered can be set by adjusting the suction regulator with reference to the conversion table we have determined. Suction must be applied during depressurization if there is an air leak of 5/min or greater coming from the patient, otherwise suction is not essential.


As the features of many brands and models of proprietaty drains arc similar, we would expect. other types could be hyperbaric compatible, but individual testing should be performed before acceptance.


英文关键词chest tube chest drain hyperbaric oxygenation flow pneumothorax pleural drainage unit suction
类型Article
语种英语
国家Australia
收录类别SCI-E
WOS记录号WOS:000236209500012
WOS关键词SYSTEMS
WOS类目Anesthesiology ; Critical Care Medicine
WOS研究方向Anesthesiology ; General & Internal Medicine
资源类型期刊论文
条目标识符http://119.78.100.177/qdio/handle/2XILL650/150676
作者单位(1)Alfred Hosp, Hyperbar Serv, Melbourne, Vic 3004, Australia
推荐引用方式
GB/T 7714
Walker, KJ,Millar, IL,Fock, A. The performance and safety of a pleural drainage unit under hyperbaric conditions[J],2006,34(1):61-67.
APA Walker, KJ,Millar, IL,&Fock, A.(2006).The performance and safety of a pleural drainage unit under hyperbaric conditions.ANAESTHESIA AND INTENSIVE CARE,34(1),61-67.
MLA Walker, KJ,et al."The performance and safety of a pleural drainage unit under hyperbaric conditions".ANAESTHESIA AND INTENSIVE CARE 34.1(2006):61-67.
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Walker, KJ]的文章
[Millar, IL]的文章
[Fock, A]的文章
百度学术
百度学术中相似的文章
[Walker, KJ]的文章
[Millar, IL]的文章
[Fock, A]的文章
必应学术
必应学术中相似的文章
[Walker, KJ]的文章
[Millar, IL]的文章
[Fock, A]的文章
相关权益政策
暂无数据
收藏/分享

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。