The performance and safety of a pleural drainage unit under hyperbaric conditions | |
Walker, KJ; Millar, IL; Fock, A | |
通讯作者 | Walker, KJ |
来源期刊 | ANAESTHESIA AND INTENSIVE CARE
![]() |
ISSN | 0310-057X |
EISSN | 1448-0271 |
出版年 | 2006 |
卷号 | 34期号:1页码:61-67 |
英文摘要 | The performance of a proprietary dry suction pleural drainage unit was measured under hyperbaric oxygenation conditions. The test pleural drainage unit was connected to pressure gauges that allowed the pressures created in the suction and collection chambers to be measured as well as the pleural drainage catheter pressures under varied suction regulator settings during compression, hyperbaric steady states and decompression. The maximum flow capacity of the unit was also measured under varying hyperbaric conditions. The Atrium Oasis Dry Suction 3600 Chest Drain brand was dramatically affected by pressure change. Nevertheless, based upon our testing, we believe it call be used safely in a hyperbaric environment provided that the following precautions are taken. Suction should not be applied during pressurization. Pressurization needs to be slow, 10 kpa/min or less. Suction is needed for air leaks of 4/min or more at pressure. At stable hyperbaric pressure, the level of suction delivered can be set by adjusting the suction regulator with reference to the conversion table we have determined. Suction must be applied during depressurization if there is an air leak of 5/min or greater coming from the patient, otherwise suction is not essential. As the features of many brands and models of proprietaty drains arc similar, we would expect. other types could be hyperbaric compatible, but individual testing should be performed before acceptance. |
英文关键词 | chest tube chest drain hyperbaric oxygenation flow pneumothorax pleural drainage unit suction |
类型 | Article |
语种 | 英语 |
国家 | Australia |
收录类别 | SCI-E |
WOS记录号 | WOS:000236209500012 |
WOS关键词 | SYSTEMS |
WOS类目 | Anesthesiology ; Critical Care Medicine |
WOS研究方向 | Anesthesiology ; General & Internal Medicine |
资源类型 | 期刊论文 |
条目标识符 | http://119.78.100.177/qdio/handle/2XILL650/150676 |
作者单位 | (1)Alfred Hosp, Hyperbar Serv, Melbourne, Vic 3004, Australia |
推荐引用方式 GB/T 7714 | Walker, KJ,Millar, IL,Fock, A. The performance and safety of a pleural drainage unit under hyperbaric conditions[J],2006,34(1):61-67. |
APA | Walker, KJ,Millar, IL,&Fock, A.(2006).The performance and safety of a pleural drainage unit under hyperbaric conditions.ANAESTHESIA AND INTENSIVE CARE,34(1),61-67. |
MLA | Walker, KJ,et al."The performance and safety of a pleural drainage unit under hyperbaric conditions".ANAESTHESIA AND INTENSIVE CARE 34.1(2006):61-67. |