Measurement of scotopic pupils. Comparison of scotopic pupil measurements using a Colvard pupilometer and the slitlamp green light test
Sekundo, E; Nietgen, G; Schmidt, JC
通讯作者Sekundo, E
来源期刊OPHTHALMOLOGE
ISSN0941-293X
出版年2003
卷号100期号:10页码:815-818
英文摘要

Purpose. The aim of the study was to compare the accuracy and the reproducibility of scotopic pupil measurements using two different methods.


Patients and methods. We developed a simple test to measure scotopic pupil diameters using a narrow slit of light through a green filter of the Haag-Streit slitlamp. A total of 100 eyes from 50 refractive surgery candidates were prospectively examined by 2 independent investigators using both the Colvard pupilometer (Oasis/USA) and Sekundo’s slitlamp green light test. Results were compared using all pairwise multiple comparison procedures (Student-Newman-Keuls method, SigmaStat/Jandel Scientific). The mean age of the patients was 36.3 years with a male:female ratio of 16:34. The colour of the iris was considered blue for 36 individuals and brown for the remainder.


Results. Reproducibility: the right eye mean pupil diameter using Sekundo’s method was 6.4 mm (+/-0.9) measured by both the first (Slit 1R) and the second (Slit 2R) investigator. The left eye mean pupil diameter was 6.5 mm (+/-1) (Slit 1L) and 6.35 mm (+/-1) (Slit 2L), respectively. The following measurements were obtained with the Colvard pupillometer: (Colv1R) = 6.25 +/- 0.85 mm, (Colv2R) = 5.99 +/- 1 mm, (Colv1L) = 6.15 +/- 0.91 mm, (Colv2L) = 6.05 +/- 0.9 mm. There were no statistically significant differences between the two investigators within the same method of examination. Comparability:The data of both investigators were combined to form 4 groups: (Slit R) = 6.4 +/- 0.96 mm, (Slit L) = 6.4 +/- 1 mm, (Colv R) = 6.05 +/- 0.9 mm, (Colv L) = 6.1 +/- 0.93 mm. There was a significant difference (p = 0.024) between the two methods for both right and left eyes. Particularly for large pupils,the slitlamp green light measurements were significantly higher (up to 0.35 mm) than those with Colvard’s device.


Conclusions. The slitlamp green light test provides similar reproducibility for the measurement of the scotopic pupil diameter as the commercially available Colvard pupilometer.


英文关键词Scotopic pupil Clovard pupillometer slitlamp green light test
类型Article
语种German
国家Germany
收录类别SCI-E
WOS记录号WOS:000186443600006
WOS关键词INFRARED PUPILLOMETER ; SIZE
WOS类目Ophthalmology
WOS研究方向Ophthalmology
资源类型期刊论文
条目标识符http://119.78.100.177/qdio/handle/2XILL650/145557
作者单位(1)Univ Marburg, Klin Augenheilkunde, D-35032 Marburg, Germany
推荐引用方式
GB/T 7714
Sekundo, E,Nietgen, G,Schmidt, JC. Measurement of scotopic pupils. Comparison of scotopic pupil measurements using a Colvard pupilometer and the slitlamp green light test[J],2003,100(10):815-818.
APA Sekundo, E,Nietgen, G,&Schmidt, JC.(2003).Measurement of scotopic pupils. Comparison of scotopic pupil measurements using a Colvard pupilometer and the slitlamp green light test.OPHTHALMOLOGE,100(10),815-818.
MLA Sekundo, E,et al."Measurement of scotopic pupils. Comparison of scotopic pupil measurements using a Colvard pupilometer and the slitlamp green light test".OPHTHALMOLOGE 100.10(2003):815-818.
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Sekundo, E]的文章
[Nietgen, G]的文章
[Schmidt, JC]的文章
百度学术
百度学术中相似的文章
[Sekundo, E]的文章
[Nietgen, G]的文章
[Schmidt, JC]的文章
必应学术
必应学术中相似的文章
[Sekundo, E]的文章
[Nietgen, G]的文章
[Schmidt, JC]的文章
相关权益政策
暂无数据
收藏/分享

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。