Comparison of the effectiveness of 2 dual-energy X-ray absorptiometers with that of total body water and computed tomography in assessing changes in body composition during weight change | |
Tylavsky, FA; Lohman, TG; Dockrell, M; Lang, T; Schoeller, DA; Wan, JY; F’uerst, T; Cauley, TA; Nevitt, M; Harris, TB | |
通讯作者 | Tylavsky, FA |
来源期刊 | AMERICAN JOURNAL OF CLINICAL NUTRITION
![]() |
ISSN | 0002-9165 |
出版年 | 2003 |
卷号 | 77期号:2页码:356-363 |
英文摘要 | Background: Little information is available on the assessment of changes in body composition as a function of weight change with the use of the fan beam of dual-energy X-ray absorptiometry (DXA). Objective: The objective was to determine the accuracy of the fan beam of the QDR 4500A densitometer and the pencil beam of the QDR 2000 densitometer in estimating changes in whole-body lean soft tissue mass (LSTMDXA) and fat mass (FM) with weight change. Design: Thirty-seven subjects who lost 5.7 +/- 4.5 kg were measured before and after weight change. Using total body water and computed tomography (CT) of the midthigh, we compared changes in FFMTBW and LSTMCT with changes in LSTMDXA. Results: Overall, compared with TBW, the fan beam gave a larger estimate of change ((x) over bar +/- SD) in LSTM (fan beam - TBW: -0.7 +/- 1.6 kg) than did the pencil beam (pencil beam - TBW: -0.1 +/- 1.6 kg)When the change in LSTM obtained with the fan beam and pencil beam was regressed against the change in FFMTBW, the slope of the line for the fan beam was 0.97 (r(2=) 0.61) and that for the pencil beam was 0.86 (r(2=) 0.61). Regression analysis showed that the results between the 2 units were not interchangeable. For the midthigh region, the change in LSTMCT. was moderately correlated with the change in LSTMDXA, with the fan beam arid pencil beam. Conclusions: The measurement of change in LSTM with the fan and pencil beams provides the same relation to changes in FFM assessed by TBW, but the 2 systems are not interchangeable. |
英文关键词 | body composition dual-energy X-ray absorptiometry DXA computed tomography weight loss lean body mass fat mass total body water |
类型 | Article |
语种 | 英语 |
国家 | USA |
收录类别 | SCI-E |
WOS记录号 | WOS:000180512200013 |
WOS关键词 | FAT-FREE MASS ; OBESE WOMEN ; CROSS-CALIBRATION ; LABELED WATER ; PENCIL-BEAM ; DEXA ; ADULTS ; VALIDATION ; HYDRATION ; VALIDITY |
WOS类目 | Nutrition & Dietetics |
WOS研究方向 | Nutrition & Dietetics |
来源机构 | University of Arizona |
资源类型 | 期刊论文 |
条目标识符 | http://119.78.100.177/qdio/handle/2XILL650/144045 |
作者单位 | (1)NIA, NIH, Bethesda, MD 20892 USA;(2)Univ Pittsburgh, Pittsburgh, PA 15260 USA;(3)SYNARC Inc, San Francisco, CA USA;(4)Univ Wisconsin, Madison, WI 53706 USA;(5)Univ Calif San Francisco, San Francisco, CA 94143 USA;(6)Univ Arizona, Tucson, AZ USA;(7)Univ Tennessee, Memphis, TN USA |
推荐引用方式 GB/T 7714 | Tylavsky, FA,Lohman, TG,Dockrell, M,et al. Comparison of the effectiveness of 2 dual-energy X-ray absorptiometers with that of total body water and computed tomography in assessing changes in body composition during weight change[J]. University of Arizona,2003,77(2):356-363. |
APA | Tylavsky, FA.,Lohman, TG.,Dockrell, M.,Lang, T.,Schoeller, DA.,...&Harris, TB.(2003).Comparison of the effectiveness of 2 dual-energy X-ray absorptiometers with that of total body water and computed tomography in assessing changes in body composition during weight change.AMERICAN JOURNAL OF CLINICAL NUTRITION,77(2),356-363. |
MLA | Tylavsky, FA,et al."Comparison of the effectiveness of 2 dual-energy X-ray absorptiometers with that of total body water and computed tomography in assessing changes in body composition during weight change".AMERICAN JOURNAL OF CLINICAL NUTRITION 77.2(2003):356-363. |
条目包含的文件 | 条目无相关文件。 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。